Operating Modes. Pneumatically diven & electronically controlled, closed, semi-closed and semi-open
• Pneumatically driven electrically controlled ventilator, can be used for emergency clinical resuscitation and respiratory therapy
Nonhazardous environmentally friendly frame, precise construction, elegant design, equipped with automated self-checking function during startup and an automatic self-calibrating sensor
Display Built-in 10.4 inch high visibility color TFT display, able-simultaneously display Pressure-Time and Frequency-Time waveforms in real time
Sigh: Ability to insert 1-8 sigh breaths in every 100 breaths, ventilation should be no less than 1.5 the tidal volume
Equipped with infrared turbine flow sensor, precisely measures gas flows, stable and reliable
• Power Failure Support Functions Automatically
converts-standby mode during
AC power outages
9 Level adjustable humidifier
• Monitor Parameters tidal volume, respiration rate, airway pressure, I E ratio, inspiratory & expiratory status, inspiratory trigger and others
• Alarms: Audible and/or visual alarms for ventilation volume, tidal volume, oxygen concentration, asphyxia, airway pressure, intubation disconnection, low oxygen, power failure and other alarms, also features a alarm
history query function
Air Compressor
• Air Compressor: 220V AC±10%, 50±1Hz, 700VA, also features good protective grounding against electrical activity
• Output Gas Pressure: 0.4MPa±20%
• Continuous output gas flow greater than or equal to 20L/min
• Instantaneous output gas flow great than or equal to 60L/min
• Equipped with professional medical grade air compressor tailored for long clinical times and low noise requirements
Model | NBV-700B (Standard Model) | NBV-700B (Advanced Model) |
Display | 10.4 inch high visibility color TFT display | |
Ventilation Mode | A/C, IPPV, SIPPV, IMV, SIMV, Standby. SNAP, MANUAL | |
Minute Volume | Greater than or equal to 18L | |
Back-up Power Supply | At least 4 hours | |
Tidal Volume | Adjustable range: 50-1500ml, Display range: 0-2000ml | |
Respiratory Rate | 1-99bpm | |
Inspiratory/expiratory (I:E) Ratio | 4:1-14 | |
PEEP | 0-10cmH2O | |
SIMV Rate | 1-20bpm | |
Inspiratory Pressure Trigger | -10-10cmH2O | |
Inspiratory Plateau | Adjustable from 0-50% of inspiratory time | |
Adjustable Oxygen Concentration | 45-100% | 21-100% |
Gas Supply Requirement | 280~600 kPa medical grade oxygen and compressed air | |
Pressure Limit | 1.0kPa~6.0kPa | |
Maximum Safe Operating Pressure | ⋜6.0kPa | |
Tidal Volume Alarm | Upper Limit: 10~2000ml, Lower Limit: 0~1800ml | |
Airway Pressure Alarm | Upper Limit: 0.1-6kPa, Lower Limit: 0-5kPa | |
Oxygen Concentration Alarm | Upper Limit: 22-100%, Lower Limit: 21-80% |