Operating Modes Pneumatically driven & electronically controlled closed semi-closed and semi-open
Pneumatically driven electrically controlled ventilator, can be used for emergency clinical resuscitation and respiratory therapy
• Display Built-in 10.4 inch color TFT fully adjustable display, able-display Pressure-Time Frequency-Time, Pressure-Volume. Frequency-Volume and any two waveforms
simultaneously on screen
• Trigger Types Pressure trigger, Flow trigger
• Sigh: Ability to insert 1-8 sigh breaths in every 100 breaths, ventilation should be no less than 15 the tidal volume
• Features built-in dual electronic flow sensors and dual respiratory modules
• Power Failure Support Functions
Automatically converts-standby mode during AC power outages
• Monitor Parameters: tidal volume, respiration rate, airway pressure, IE ratio, inspiratory & expiratory status, inspiratory tigger and others
• Equipped with infrared turbine flow sensor precisely measures gas flows, stable and reliable
.9 Level adjustable humidifier
• Nonhazardous environmentally friendly frame, precise construction, elegant design, equipped with automated self-checking function during startup and an automatic self-calibrating sensor
• Alarms Audible and/or visual alarms for ventilation volume, tidal volume, oxygen concentration, asphyxia, airway pressure, intubation disconnection, low oxygen, power failure and other alarms, also features a alarm history query function
Air Compressor
• Air Compressor: 220V AC±10%, 50±1Hz 700VA, also features good protective grounding against electrical activity
• Output Gas Pressure: 0.4MPa±20%
• Continuous output gas flow greater than or equal to 20L/min
• Instantaneous output gas flow great than or equal to 60L/min
Model | NBV-900B (Standard Model) | NBV-900B (Advanced Model) |
Display | 10.4 inch high visibility color TFT display | |
Minute Volume | Greater than or equal to 18L | |
Ventilation Mode | A/C, SIPPV, IPPV, SIMV, PSIMV, PSV, CPAP, PCV, VCV SNAP, MANUAL | |
Back-up Power Supply | At least 4 hours | |
Tidal Volume | Adjustable range: 20-1500ml, Display range: 0-2000ml | |
Respiratory Rate | 1-99bpm | |
Inspiratory/expiratory (I:E) ratio | 4:1-1:4 | |
PEEP | 0-10cmH2O | |
SIMV Rate | 1-20bpm | |
Inspiratory Pressure Trigger | -10-10cmH2O | |
Trigger Sensitivity (flow) | 1-10L/m | |
Inspiratory Plateau | Adjustable from 0-50% of inspiratory time | |
Adjustable Oxygen Concentration | 45-100% | 21-100% |
Gas Supply Requirement | 280~600 kPa medical grade oxygen and compressed air | |
Pressure Limit | 1.0kPa~6.0kPa | |
Maximum Safe Operating Pressure | ≤6.0kPa | |
Tidal Volume Alarm | Upper Limit: 10~2000ml, Lower Limit: 0~1800ml | |
Airway Pressure Alarm | Upper Limit: 0.1~6kPa, Lower Limit: 0-5kPa | |
Oxygen Concentration Alarm | Upper Limit: 22~100%, Lower Limit: 21~80% |